Misplaced Pages

Ampere balance

Article snapshot taken from Wikipedia with creative commons attribution-sharealike license. Give it a read and then ask your questions in the chat. We can research this topic together.
This article relies largely or entirely on a single source. Relevant discussion may be found on the talk page. Please help improve this article by introducing citations to additional sources.
Find sources: "Ampere balance" – news · newspapers · books · scholar · JSTOR (February 2011)
This article needs to be updated. The reason given is: it needs to reflect the 2019 revision of the SI, which came into effect on May 20, 2019. Please help update this article to reflect recent events or newly available information. (February 2020)
(left) Rosa-Dorsey type Ampere balance at the US National Bureau of Standards (now NIST) in 1927. The current coils are visible under the balance, attached to the right balance arm.
(right) Ampere balance constructed at the former National Institute of Advanced Industrial Science and Technology, Japan, now in the National Museum of Nature and Science, Tokyo, Japan.
The ampere balance

The ampere balance (also current balance or Kelvin balance) is an electromechanical apparatus used for the precise measurement of the SI unit of electric current, the ampere. It was invented by William Thomson, 1st Baron Kelvin.

The current to be measured is passed in series through two coils of wire, one of which is attached to one arm of a sensitive balance. The magnetic force between the two coils is measured by the amount of weight needed on the other arm of the balance to keep it in equilibrium. This is used to calculate the numerical value of the current.

The main weakness of the ampere balance is that the calculation of the current involves the dimensions of the coils. So the accuracy of the current measurement is limited by the accuracy with which the coils can be measured, and their mechanical rigidity.

A more complicated version of an ampere balance, that removes this source of inaccuracy by a calibration step, is the Kibble balance, invented by Bryan Kibble in 1975. This experimental device was developed at government metrology laboratories worldwide with the goal of providing a more accurate definition of the kilogram, the world's standard of mass. In this application, the Kibble balance functions in the reverse sense to the Ampere balance: it is used to weigh the International Prototype of the Kilogram, defining the kilogram in terms of an electric current and a voltage. Since voltage and current themselves are already defined in terms of fundamental physical constants like the Planck constant and the speed of light, this has created a new definition of the kilogram in terms of these fundamental constants. This is a more stable definition than the previous one, which was based on the physical prototype kilogram which is vulnerable to damage and deterioration over the years.

Usage

Slider of Kelvin's ampere balance

Approximate readings may be obtained by reading the position of the weight on the scale, or a more accurate reading may be obtained as follows: The upper edge of the shelf on which the weights slide is graduated into equal divisions, and the weight is provided with a sharp tongue of metal in order that its position on the shelf may be accurately determined. Since the current passing through the balance when equilibrium is obtained with a given weight is proportional to the square root of the couple due to this weight, it follows that the current strength when equilibrium is obtained is proportional to the product of the square root of the weight used and the displacement of this weight from its zero position. Each instrument is accompanied by a pair of weights and by a square root table, so that the product of the square roots of the number corresponding to the position of the sliding weight and the ascertained constant for each weight, gives at once the value of the current in amperes. Each of these balances is made to cover a certain range of reading. Thus, the centiampere balance ranges from 1 to 100 centiamperes, the deciampere balance from 1 to 100 deciamperes, the ampere balance from 1 to 100 amperes, etc.

References

Categories: