Misplaced Pages

Amplitude integrated electroencephalography

Article snapshot taken from Wikipedia with creative commons attribution-sharealike license. Give it a read and then ask your questions in the chat. We can research this topic together.
(Redirected from Cerebral function monitoring) Technique for monitoring brain function in intensive care settings Medical diagnostic method
Amplitude integrated electroencephalography
A normal CFM trace in a term infant who is several days old. The amplitude-integrated trace is in the top half of the screen displaying left- and right-sided traces. There is a normal baseline and upper limit, sleep-wake cycling, and no seizures.
Purposemonitoring brain function in intensive care

Amplitude integrated electroencephalography (aEEG), cerebral function monitoring (CFM) or continuous electroencephalogram (CEEG) is a technique for monitoring brain function in intensive care settings over longer periods of time than the traditional electroencephalogram (EEG), typically hours to days. By placing electrodes on the scalp of the patient, a trace of electrical activity is produced which is then displayed on a semilogarithmic graph of peak-to-peak amplitude over time; amplitude is logarithmic and time is linear. In this way, trends in electrical activity in the cerebral cortex can be interpreted to inform on events such as seizures or suppressed brain activity. aEEG is useful especially in neonatology where it can be used to aid in diagnosis of hypoxic ischemic encephalopathy (HIE), and to monitor and diagnose seizure activity.

Interpretation of the aEEG

The CFM readout offers an integrated trace in one pane and a non-integrated trace in another pane (see image). Modern machines give a readout for each hemisphere corresponding to the positions of electrodes placed on the patient's head. The characteristics of the CFM include the 'baseline' which should be more than 5 μV, the upper limit of the trace which should be more than 10 μV, and the presence of 'sleep wake cycling' whereby the trace is expected to narrow and broaden over time. Seizures appear on the trace as regions of high activity with a raised and compacted trace in the aEEG pane; this would correspond to high-amplitude, repetitive waveforms in the non-integrated pane. A low-amplitude or 'suppressed' trace is prognostically concerning as it indicates abnormally low brain activity. A further possible pattern is a 'burst suppression' trace which consists of a low-amplitude signal interspersed with periods of high activity on the aEEG readout. This also carries a poor prognosis.

See also

References

  1. Maynard D, Prior P, Scott D. Device for continuous monitoring of cerebral activity in resuscitated patients. Br Med J. 1969 Nov 29;4(5682):545-6. PMID 5354856
  2. Azzopardi D. TOBY Cerebral Function Monitoring: Addition to CFM handbook for users of the Olympic CFM 6000. Imperial College London, 2004.
Seizures and epilepsy
Basics
Management
Personal issues
Seizure types
Focal
Seizures
Simple partial
Complex partial
Gelastic seizure
Epilepsy
Temporal lobe epilepsy
Frontal lobe epilepsy
Rolandic epilepsy
Sleep-related hypermotor epilepsy
Panayiotopoulos syndrome
Vertiginous epilepsy
Generalised
Status epilepticus
Myoclonic epilepsy
Related disorders
Organizations
Emerging technologies
Fields
Biomedical
Topics
Electroencephalography (EEG)
Related tests
Evoked potentials
Negativity
Bereitschaftspotential
ELAN
N100
Visual N1
N170
N200
N2pc
N400
Contingent negative variation (CNV)
Mismatch negativity
Positivity
C1 & P1
P50
P200
P300
P3a
P3b
P600 (late positivity)
Late positive component
Neural oscillations
Topics
Categories: